industrial spectroradiometer iSR900
The industrial spectroradiometer iSR900 is an autonomous and high-resolution measuring system for fast and precise measurements in the UV, visible spectral range and IR.
The spectroradiometer consists of an array spectrometer without moving parts and a silicon detector with 2048 pixels. The spectral resolution is 0.44 nm.
The iSR900 is combined with an SMA-905 quartz light guide and our radiometric measuring heads for irradiance and illuminance measurement.
The iSR900 is factory calibrated to PTB. A DAkkS test in our accredited laboratory is optionally available. The iSR900 thus allows accurate spectroradiometric measurements for the evaluation of irradiance and illuminance. The low-scatter design achieves high sensitivity in the UV range. Control and data evaluation take place in the iSR900 so that only a USB or RS485 / RS232 interface is required for the evaluation.
With trigger input and output the iSR900 is ideal for automated measurements.
HIGHLIGHS OF Industrual SPECTRODATIOMETER iSR900
- Spectral range 200-1100 nm
- Nearly scatter-free grid with 300 lines/mm
- Spectral resolution 0.4 nm and 2,3 nm bandwidth
- Order filtering
- Radiometric and effect-related measurements
- USB, RS485 or RS232
iSR900 Sample commands
iSRSerialNr? | Serial number query |
iSRFirmware? | Queryfirmware version |
iSRReset! | Reset command |
iSRIntTime: 400! | Setting the integration time |
iSRMeasAVG?! | Request/Command Averages |
iSRCalibDate?: | Calibration date request |
iSRStartDark! | Start dark measurement |
iSRStartMeas! | Start measurement command |
iSRMeasResult? | Inquiry of measurement result |
TECHNICAL DATA Industrial SPECTRORADIOMETER iSR900
Spectral range | 200 - 1100 nm |
Spectral resolution | 0,44 nm (pixel pitch) |
Spectral bandwidth | 2,3 nm (FWHM) |
Grating | 300 l/mm, holografic |
Focal length | 75 mm |
Blaze wavelength | 250 nm |
Detector | SI |
Pixel size | 14 x 200 µm |
Number of pixel | 2048 |
Resolution | 16 bit |
Integrations time | 0,1 ms to 60 s |
measurement duration, typ. | 300 ms |
Interface | USB, RS232 or RS485 |
Trigger | I/ O, rear connection |
Dimensions | 250 x 320 x 100 mm |
Gewicht | 3,5 kg |
Supply voltage | 230 V, 50 Hz |
Power | < 25 W |
Operating temperature | 5 to 40 °C |
Storage temperature | -10 to 60 °C |
Humidity | <80%, non-condensating |
Communication | ASCII with CRC-16 |