ISO 17025 Calibration of Spectroradiometers and Spectrometers
The calibration laboratory of Opsytec Dr. Gröbel GmbH calibrates spectroradiometers and spectrometers under accreditation to DIN EN ISO/IEC 17025 (D-K-20284-01-00). The spectral irradiance responsivity is calibrated according to ASTM G138 in the spectral range from 200 nm to 2400 nm, with an expanded measurement uncertainty from 2.5 %. You receive an ISO 17025 calibration certificate that is traceable to the PTB and internationally recognised through the EA and ILAC agreements – for Opsytec instruments as well as for spectroradiometers from other manufacturers.
What the calibration determines
A spectrometer delivers count rates (counts) that depend on the detector, the integration time and the input optics. The calibration assigns to each wavelength how many counts per millisecond a spectral irradiance of 1 W/(m²·nm) produces. Only then does the spectrometer become a spectroradiometer that delivers absolute values: spectral irradiance and, derived from it, irradiance per spectral band, illuminance, PPFD or weighted quantities.
The complete measuring chain – spectrometer, light guide and input optics – is calibrated at a defined instrument setting. The calibration certificate applies to exactly this configuration.
Accredited scope
The expanded measurement uncertainty depends on the spectral range:
| Spectral range | Expanded measurement uncertainty |
|---|---|
| 200 nm to 230 nm | 4.0 % |
| 230 nm to 280 nm | 3.0 % |
| 280 nm to 315 nm | 2.8 % |
| 315 nm to 400 nm | 2.6 % |
| 400 nm to 850 nm | 2.5 % |
| 850 nm to 1000 nm | 3.5 % |
| 1000 nm to 1400 nm | 4.0 % |
| 1400 nm to 1450 nm | 5.5 % |
| 1450 nm to 2000 nm | 5.0 % |
| 2000 nm to 2400 nm | 10 % |
The binding reference is the annex to the accreditation certificate of the calibration laboratory (valid from 04.11.2026).
Factory calibration or ISO 17025 calibration
Every Opsytec spectroradiometer is supplied factory-calibrated and traceable to the PTB. For routine measurements in your own process, this is often sufficient. You need an ISO 17025 calibration when traceability has to be demonstrated to third parties: for reference instruments against which other measuring devices are adjusted, for audits under IATF 16949 or in a GMP environment, and in laboratories that themselves work to ISO/IEC 17025.
In accordance with EA-3/01, please state clearly when ordering whether you require a factory calibration or an ISO 17025 accredited calibration.
Calibration procedure
- Record the instrument, input optics, light guide and required instrument setting.
- Check and document the incoming condition and function.
- Record the spectrum of an irradiance standard traceable to the PTB and subtract the dark signal.
- Calculate the spectral irradiance responsivity for each wavelength range and determine the measurement uncertainty.
- Issue the calibration certificate with result, measurement conditions and measurement uncertainty, and store the new calibration in the instrument or software.
Which instruments are calibrated
We calibrate the Opsytec spectroradiometers UVpad, UVpad E, SR900 and ISR900 Pro as well as spectroradiometers and array spectrometers from other manufacturers. For third-party instruments, we clarify in advance – based on manufacturer, type, input optics, spectral range and software – how the calibration is transferred to the instrument. More on this under Calibration of instruments from other manufacturers.
Lamp and radiometer as a set
Also new in the accredited scope: where a radiometer always measures at the same source, we calibrate both together – the spectral irradiance of the source and the radiometer at it, in the spectral range from 200 nm to 1000 nm. The spectral mismatch between calibration lamp and your source is eliminated. Details are given under UV calibration for radiometers and UV sensors.
Frequently asked questions about spectroradiometer calibration
How does the ISO 17025 calibration differ from the previous testing?
Until November 2026, the factory calibration of a spectroradiometer was tested in our accredited testing laboratory. Since then, the calibration itself has been accredited: you receive an ISO 17025 calibration certificate with the measurement uncertainty for each spectral range, which serves directly as proof of traceability.
Which spectral range is accredited?
The accredited range extends from 200 nm to 2400 nm. The expanded measurement uncertainty lies between 2.5 % in the visible range and 10 % above 2000 nm. We calibrate the range that your instrument actually covers.
Are spectroradiometers from other manufacturers also calibrated?
Yes. This requires input optics for irradiance and a way of storing the calibration in the instrument or software. We clarify this in advance based on manufacturer, type and software.
How often should a spectroradiometer be calibrated?
Twelve months is a practical starting point, to be confirmed or shortened based on drift and operating conditions. After replacing the light guide or input optics, after mechanical shocks or in case of doubt, a new calibration is necessary, because the calibration certificate only applies to the calibrated measuring chain.